3. Samples stored for a long period of time are used as specimens Sample can be stored for upto a week at 2–8°C. Only serum must be
3. Anti-DNP may be found in diseases other than
SLE. Low titers have been detected in rheumatoid
arthritis, chronic hepatitis, periarteritis nodosa,
dermatomyositis, scleroderma, atypical pneumonia,
tuberculosis and lymphoma.
Troubleshooting
Problem: False positive results
Possible causes Solutions
1. Contamination of the latex reagent with positive control or positive
sample
Precaution should be taken so that the dropper tip of the reagent does
not touch the samples and controls on the glass slide
2. Samples are stored for a long period Should a delay in testing occur, store samples at 2–8°C. Samples can
be stored for upto a week at 2–8°C
3. Cross contamination due to the usage of the same mixing stick Separate mixing stick should be used for mixing the controls and the
sample
4. Markedly lipemic, hemolyzed and contaminated serum samples Using lipemic, hemolyzed and contaminated samples produce nonspecific results. Avoid using such samples
5. Drying of the reagent on the slide Do not read results beyond 2 minutes. The test should not be carried
out directly under the fan
6. Presence of dust or debris on the glass slide used Dust or debris could be misinterpreted as agglutination therefore only
can dry glass slides must be used for testing
7. Latex particles contaminated with positive control/positive sample Care must be taken to see that the latex reagent dropper tip does not
touch the sample or control taken on the slide during dispensing of
the reagen
8. Wrong dropper used for dispensing the sample Accessories provided with the kit only must be used for optimum
results
9. Increase in drop size, thereby leading to excess reagent dispensed Excess reagent dispensed gives false positive results at borderline
concentrations. Ensure that exactly one drop of reagent is dispensed
onto the slide
10. Reagent dropper not held vertically while dispensing To ensure accurate dispensing of the reagent, hold the reagent dropper vertically while dispensing the reagent
11. Cross contamination due to the usage of the same mixing stick Separate mixing stick should be used for mixing the controls and the
sample
12. Dried latex particles observed in the latex reagent:
• During slide test with negative control
• In the dropper of the vial (due to freezing of the latex reagent
during storage)
• Improper dispensing of the entire reagent from dropper
Immediately after performing the test, transfer the contents of the
reagent dropper back into the reagent vial
Ensure that no reagent is left behind in the dropper
Close the cap of the reagent vial properly and store it back at 2–8°C.
Do not freeze the reagent vial
13. Low titers of Anti-DNP may also be found in clinical conditions
such as RA, chronic hepatitis, periarteritis nodosa, dermato myosis, Scleroderma, atypical pneumonia, tuberculosis and lymphoma.
The clinical history of the patient should be checked for any of these
disorders
662 Concise Book of Medical Laboratory Technology: Methods and Interpretations Problem: Delayed positive results
Possible causes Solutions
1. The positive control may have deteriorated due to contamination
or exposure to extreme temperatures
Check the performance of the latex reagent; using known positive
samples, if the latex reagent is working then the positive control may
have deteriorated
Possible causes Solutions
1. Reagents not brought to room temperature before testinG Bring the reagents to room temperature before carrying out the test
Possible causes Solutions
1. Cold reagents have been tested The reagents should be tested only after attaining room temperature
Problem: False negative results
Possible causes Solutions
1. The reagent may be damaged due to microbial contamination or
exposure to extreme temperatures
Performance of the reagents can be verified by using positive control/
known positive sample
2. Weak agglutination may be interpreted as negative Shake the latex reagent well before use to disperse the latex particles
uniformly and improve test readability
3. Samples stored for a long period of time are used as specimens Sample can be stored for upto a week at 2–8°C. Only serum must be
used for testing
4. Sample may be hemolyzed or contaminated Avoid using hemolyzed or contaminated samples for testing
5. If the conclusion of false negative results has been arrived at by
comparison with another kit, this other kit could be giving a false
positive reaction
Run the test with a third kit to validate results
6. The latex reagent might have been frozen The latex reagent should never be frozen as freezing leads to the dissociation of the DNP coated on the latex. The free IgG neutralizes the
RF present in the sample thereby leading to false negative results
Problem: Positive control giving negative reaction
Problem: Delayed agglutination
AUSTRALIA ANTIGEN HBSAG (VIRUTEX HBsAg)
(Courtesy: Tulip Group of Companies)
Slide Test for Hepatitis B Surface Antigen
Summary
No comments:
Post a Comment
اكتب تعليق حول الموضوع